Home » Climatic Chambers » Climatic Chambers » Climatic Chambers CH 150 and CH 250
These climatic chambers are designed for precise control of temperature and humidity. They are ideal for various laboratory applications that require controlled environmental conditions.
Klikom na naziv možete pogledati detaljne informacije za svaki model.
CLIMATIC CHAMBER CH 150: Capacity 150 liters, working temperature range from 0 ÷ 85 °C (without humidity) and 10 ÷ 70 °C (with humidity), humidity fluctuation ± 3% (10 °C ÷ 70 °C / 55 ÷ 95% RH). External dimensions 690 x 1520 x 790 mm, weight 145 kg.
CLIMATIC CHAMBER CH 250: Capacity 250 liters, working temperature range from 0 ÷ 85 °C (without humidity) and 10 ÷ 70 °C (with humidity), humidity fluctuation ± 3% (10 °C ÷ 70 °C / 55 ÷ 95% RH). External dimensions 740 x 1680 x 885 mm, weight 185 kg.
Both chambers come with a digital LCD display, mini printer, three standard shelves, and safety class 3.1 (DIN 12880).
| Feature | CH 150 | CH 250 |
|---|---|---|
| Power supply | 220 ±10% V | 220 ±10% V |
| Capacity | 150 L | 250 L |
| Maximum shelf load | 10 kg | 10 kg |
| Safety class | 3.1 | 3.1 |
| External dimensions | 690 x 1520 x 790 mm | 740 x 1680 x 885 mm |
| Internal dimensions | 550 x 405 x 670 mm | 600 x 500 x 830 mm |
| Display | Digital backlit LCD | Digital backlit LCD |
| Humidity fluctuation | ± 3% (10…70°C / 55…95% RH) | ± 3% (10…70°C / 55…95% RH) |
| Frequency | 50/60 Hz | 50/60 Hz |
| Warranty | 2 years | 2 years |
| Working method | Forced ventilation | Forced ventilation |
| Number of shelves | 3 | 3 |
| Homogeneity in space (with humidity) | ±1.5 °C (+10…+70 °C / 55…95 %RH) | ±1.5 °C (+10…+70 °C / 55…95 %RH) |
| Homogeneity in space (without humidity) | ±0.5 °C | ±0.5 °C |
| Weight | 145 kg | 185 kg |
| Absorbed power | 2200 W | 2200 W |
| Programs | Customisable programs with ramps from 1 to 100 steps and Fix | Customisable programs with ramps from 1 to 100 steps and Fix |
| Coolant quantity | 280 g | 280 g |
| Coolant type | R1234ZE | R1234ZE |
| Resolution | 0.01 °C / 0.1% RH | 0.01 °C / 0.1% RH |
| Data transfer | Numerical or graphic via printer receipt | Numerical or graphic via printer receipt |
| Stability over time without humidity | ±0.2 °C | ±0.2 °C |
| Working temperature | +10…+70 °C (with humidity) / 0…+85 °C (without humidity) | +10…+70 °C (with humidity) / 0…+85 °C (without humidity) |
| Timer | (99:59 and Fix) digital with date and time for GLP functions | (99:59 and Fix) digital with date and time for GLP functions |
| Temperature variation over time in 1 point (with humidity) | ±0.5 °C | ±0.5 °C |
| Useful volume | 150 L | 250 L |