Home » Multiparameters » Portable » XS PC 80 PRO Stirrer – Bench Multiparameter
Digital benchtop multiparameter meter for advanced laboratory analysis.
Multiview screen (up to 6 parameters can be displayed simultaneously at user’s choice).
Automatic or manual 5-point pH calibration with automatic standard recognition.
ISE calibration from 2 to 5 points.
Automatic or manual ORP calibration on 1 point.
Automatic EC calibration up to 1 to 4 points with recognition of standards and calculation of different constants of
cell + 1 point manual.
Automatic scale change, automatic and manual temperature compensation (0…100.0 °C). Reference temperature (15…30 °C).
Linear (0.00…10.00%) and not linear temperature coefficient for ultrapure water.
Compatible with 4-ring and 2-ring cells, with indication of selected constant.
Calibration expiration and/or measurement lockout warning.
Alarm setting and display functions in analog format, “Stop-reading,” stability indication, self-diagnostic messages.
Advanced Management of Users (Admin, Manager, User, Guest) with different possibilities of access to instrumental functions.
Different data logger modes (Manual, Automatic, Min/Max, Delta value).
Analysis Menu for a complete edit of your measurements: you can enter all useful information about the sample and the test.
Customizable print templates. Multilingual Setup menu (Ita – Eng – Deu – Fra – Esp – Por).
Free Data-Link+ software at www.giorgiobormac.com for PC-based data management, including Audit Trail.
2 USB outputs for external keyboard and PC connection. RS232 port for printer connection.
Measuring ranges:
pH: -2,000…20,000 (0.001 pH)
mV: -2000…+2000 (0.1/1 mV)
ORP: -2000…+2000 (0.1/1 mV)
ISE: 0.001…19999 ppm
Cond: 0.0 µS …1000 mS (0.1 µS – automatic)
Resistivity: 1 Ohm …10 MOhm
Salinity: 0.01…100.0 ppt
TDS: 0.1 mg/l…500.0 g/l
Temperature: -30…130.0 °C (0.1 °F)
Supplied as standard:
Magnetic stirrer, S7/BNC cable, NT 55 temperature probe, electrode holder stand, colored buffer solutions, external keypad, USB cable and multi socket power supply.
Variations:
Variations:
With XS Standard DHS pH electrode and VPT1 conductivity cell.
With VPT1 conductivity cell, without pH electrode.
With XS Standard S7 pH electrode, without conductivity cell.
Without pH electrode and conductivity cell.
With EPT4 conductivity cell and Standard S7 electrode.
With EPT4 conductivity cell, without pH electrode.
With EPT4 conductivity cell and Standard DHS electrode.
| Features | Values |
|---|---|
| Accuracy (Cond) | ±0.5 % f.s. |
| ISE accuracy | 0.5 % f. s. for monovalent ions – 1 % for divalent ions |
| Accuracy (pH) | ±0.002 pH |
| Accuracy (Salinity) | ±1 % f.s. |
| Accuracy (TDS) | ±1 % f.s. |
| Accuracy (Temperature) | ±0.2 °C |
| Magnetic stirrer | Yes, independently controlled |
| Power supply | 100 – 240 V |
| Temperature calibration | Yes |
| AQA calibration | Yes |
| Temperature coefficient (max) | 0.00 … 10.00 % / °C |
| DHS compatibility | Yes |
| Temperature compensation | 0.0 … +100.0 °C |
| Ultrapure water compensation | Yes for values below 10 µS |
| PC Communication | USB type B |
| Printer connection | RS 232 |
| Stirring speed control | 15 levels |
| Cell constant | 2 rings: 0.1 / 1 / 10 | 4 rings: 0.55 |
| Criteria for stability of the measure | Med – High – Tit | Seconds (ISE) |
| Dimension | 324 x 185 x 56 mm |
| Display | 4.3″ high-definition, backlit color TFT |
| TDS factor | 0.40 … 1.00 |
| Warranty | 3 years |
| AQA | Yes |
| Calibration data indication | Yes with reports |
| Buffer used indication | Yes with icon |
| Electrode status indication | Yes with graph and icon |
| Inputs | 2 x BNC – 2 x RCA – Multipin – 4 mm banana |
| Memory | 8000 |
| Name | PC 80 Pro |
| Parameters | pH – mV – ORP – ISE – Conductivity – TDS – Salinity – Resistivity – Temperature |
| Password | 6 to 10 characters with a capital letter, a number and a symbol |
| Weight | 1.255 kg |
| Calibration points (Cond) | 1 … 4 automatic values – 1 user value |
| Calibration points (ION) | 2 … 5 |
| Calibration points (ORP) | 1 automatic value – 1 user value |
| Calibration points (pH) | 1 … 5 automatic values – 1 … 5 user values |
| Measuring range (Cond) | 0.00 µS … 1000 mS |
| Measuring range (ION) | 0.001 … 19999 mg/l |
| Measuring range (mV ORP) | -2000 … +2000 mV |
| Measuring range (pH) | -2.000 … +20.000 pH |
| Measuring range (Resistivity) | 1.00 O … 10.00 MO |
| Measuring range (Salinity) | 0.01 ppm … 100 ppt |
| Measuring range (TDS) | 0.1 mg/l … 500 g/l |
| Measuring range (Temperature) | -30.0 … +130.0 °C |
| Calibration Report | Yes, printable |
| Resolution (Cond) | 0.01 / 0.1 / 1 (automatic scale) µS/cm – mS/cm |
| Resolution (ORP) | 0.1 / 1 (automatic) mV |
| Resolution (pH) | 0.1 / 0.01 / 0.001 pH |
| Resolution (Temperature) | 0.1 °C |
| Buzzer | Yes |
| pH alarm range | Min – Max |
| Recognized Cond buffers | 84 µS – 147 µS – 1413 µS – 12.88 mS – 111.8 mS |
| Recognized pH buffers | USA – NIST – DIN 19266 – Mettler Toledo Eu – Merck |
| Working temperature | 0 … +45 °C |
| Reference temperature | +15 … +30 °C |
| Timer calibration due with alarm | Yes |
| Cell type | 2 rings – 4 rings |
| ION measuring unit | mg/l – g/l – mol/l |
| Stirring speed | 0 … 3000 rpm |
| Analog view | Yes |
| Max stirring quantity | 2 L |