PT Series Pipette Checking

Complete system for pipette checking. Pipettes are widely used instruments in laboratories and their proper maintenance is essential to ensure correct and reliable results. Each laboratory should establish a regular verification program to carry out preventive maintenance or (re)calibration if necessary.

This series of balances is designed for the verification of piston pipettes using the gravimetric method. The embedded “Pipette Check” software allows you to test the correct operation of a pipette by selecting a verification method. At the end, a detailed report is generated with the overall test result (compliant or not-compliant pipette) and the relative partial values.

The -PT models include an evaporation trap that consists of a special weighing container designed to significantly reduce the evaporation of the liquid which occurs naturally during the pipette test procedure. If not well taken into consideration, this effect will negatively affect the tests by increasing the measurement error, leading to an incorrect final result, especially in the case of small volumes. The evaporation trap creates a saturated area of water vapor above the test tube containing the liquid from the pipette, thus minimizing evaporation during the procedure and obtaining reliable results.

  • Touchscreen display
  • Internal database
  • GLP setup
Model Capacity Resolution Calibration Pan Size Verification
PT124 120g 0,1mg Automatic internal Ø 80 mm
PT214 220g 0,1mg Automatic internal Ø 80 mm
PT314 310g 0,1mg Automatic internal Ø 80 mm

Data sheet

<

Column Height 360 mm
Display Type GRAPHIC
Tare Range (subtractive) up to balance capacity
Operating Temperature Range 5 – 35°C (15-30°C for verified models)
Power Supply 110-230Vac. 50/60Hz, output 24V 500mA 13VA
Max Power Consumption 2 W
Calibrates With (Class) 200 g (E2)
Response Time (sec.) ≤4
Linearity ± 0,0003g
Repeatability

Check other products

Ask us anything

We will answer a soon as posible.