The Crystal Clinic water purification system is designed to provide purified water for smooth and reliable operation of automated biochemistry analyzers.
Crystal Clinic has two reverse osmosis modules and a high-capacity deionization cartridge (DI). Large DI cartridge volume significantly reduces running costs.
Crystal Clinic has two water quality sensors. The first sensor controls the quality of water in the storage tank. The second sensor controls the quality of water supplied to a biochemical analyzer.
The Crystal Clinic water purification system produces pure water that complies with ISO 3696 Grade II water requirements.
If the feed water does not meet specifications, additional pre-treatment solutions are available.
| Parameter | Clinic |
|---|---|
| Catalogue No. | CL-2910 |
| Grade 2 water resistivity at 25 °C | >10 MΩ x cm |
| Grade 2 water conductivity at 25 °C | < 0.1 µS/cm |
| TOC level | < 30 ppb |
| Particles >0.22 μm | < 1/mL |
| Nominal flow, pure water (to storage tank) | 20 L/h |
| Dispense flow to analyzer | 4 L/h |
| Dispense pressure (adjustable) | 100–180 kPa |
| Volumetric dispense | – |
| Dimensions (WxDxH), cm | 40x50x55 cm |
| System weight, kg | 26 kg |
| Operating weight, kg | 39 kg |
| Part number | Description | Comments |
|---|---|---|
| 10410 | Replacement pre-filter module | |
| 10113 | High-capacity deionization module | |
| 10013 | Replacement 0.22 µm dispense microfilter, sterile |
| Part number | Description | Comments |
|---|---|---|
| 10112 | Pressurized water storage tank, 60 L | |
| 10114 | Pressurized water storage tank, 100 L | |
| 10017 | Replacement sterilization UV bulb for 24V configuration | |
| 410226 | AquaStop | Water leakage protection system |
| 410143 | DQ/IQ/OQ/PQ documentation |
We will answer a soon as posible.